Proficiency testing for Thiocyanate in water from a Coke plant

Water
Chemicals
Research and Innovation
Type:
Report
Area:
Research and Innovation
Series:
Open reports (Series D)
Release year:
2014
Number of pages:
12

300 kr

Download for free

2014. 12 p. (Jernkontorets forskning D 859)

A proficiency test on determination of thiocyanate in water from a coking plant has been carried out. Nine laboratories from six European countries participated using their normal analytical methods. The standard deviation between laboratories for thiocyanate levels of 300-320 ppm was about 35 ppm (11-12 % relative standard deviation) which seems acceptable given that several different methods were used.